2015 New Analytical Formulae to Calibrate HPGe Well-type Detectors Efficiency and to Calculate Summing Corrections
M.I. Abbas
Author Affiliations +
J. Phys. Math. 6(1): 1-6 (2015). DOI: 10.4172/2090-0902.1000134

Abstract

Direct mathematical method is used in this paper to calculate the photo peak efficiency of a system containing a radioactive source placed in the well of a well type detector. Attenuation of the source and detector’s walls are included. In addition summing corrections are also included. The comparison between published experimental values and present calculated values are given. This work proved quite success in predicting the photo peak efficiency of certain detectors even for a radionuclide emitting rays with different energies.

Citation

Download Citation

M.I. Abbas. "New Analytical Formulae to Calibrate HPGe Well-type Detectors Efficiency and to Calculate Summing Corrections." J. Phys. Math. 6 (1) 1 - 6, 2015. https://doi.org/10.4172/2090-0902.1000134

Information

Published: 2015
First available in Project Euclid: 23 July 2015

zbMATH: 1326.81282
Digital Object Identifier: 10.4172/2090-0902.1000134

Keywords: Azimuth angle , Detectors , Photo peak efficiency , Photons